6 List Sequence Measurements
List Sequence Measurements
Parameter Type
String
Mode
Sequence Analyzer
Remote
Command
:SERVice[:PRODuction]:LSEQuencer:ADVanced:TDSCdma:LBER:PATTern :LOAD
<String>
:SERVice[:PRODuction]:LSEQuencer:ADVanced:TDSCdma:LBER:PATTern :LOAD?
Example
:SERV:LSEQ:ADV:TDSC:LBER:PATT:LOAD “
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF”
:SERV:LSEQ:ADV:TDSC:LBER:PATT:LOAD?
Dependencies
The characters of the string must be ‘0’~’9’, ‘A’~’F’ or ‘a~f’. The length of string must be 61. Each character
of the string stands for four bits.
Preset
“FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF”
Force Restart
Yes
State Saved
Saved in instrument state.
Initial S/W Revision
A.12.00
Help Map ID
0
Loopback BER CodeRMC(TD-SCDMA)
This parameter specifies the RMC(Reference Measurement Code) decoder type used in TD-SCDMA
Loopback BER calculation. Its type is bool. TRUE means it is single code and FALSE means it is multi code.
parameter_table_23.52941
52.94118
Key Path
SCPI Only
Mode
Sequence Analyzer
Remote Command
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:RMC:SCODe TRUE | FALSE
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:RMC:SCODe ?
Example
:LSEQ:ADV:TDSC:LBER:RMC:SCOD true
: LSEQ:ADV:TDSC:LBER:RMC:SCOD ?
State Saved
Saved in instrument state.
Initial S/W Revision
A.12.00
Help Map ID
0
Loopback BER CodeLevel(TD-SCDMA)
This parameter specifies the code level of the Reference Measurement Code used in TD-SCDMA Loopback
BER calculation.
parameter_table_23.52941
52.94118
Key Path
SCPI Only
Sequence Analyzer User's & Programmer's Reference
301