![Keysight E6650A EXF User Programming Manual Download Page 79](http://html1.mh-extra.com/html/keysight/e6650a-exf/e6650a-exf_user-programming-manual_1961647079.webp)
2 Programming the Test Set
STATus Subsystem
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Positive Transition
This command determines which bits in the Questionable Integrity Condition register will set the
corresponding bit in the Questionable Integrity Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:PTRansition <integer>
:STATus:QUEStionable:INTegrity:PTRansition?
Example
STAT:QUES:INT:PTR 8 Measurement ‘became uncalibrated’ Summary will be reported to the
Integrity Summary of the Status Questionable register.
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Signal Register
Questionable Integrity Signal Condition
Questionable Integrity Signal Enable
Questionable Integrity Signal Event Query
Questionable Integrity Signal Negative Transition
Questionable Integrity Signal Positive Transition
Questionable Integrity Uncalibrated Register
Questionable Integrity Uncalibrated Condition
Questionable Integrity Uncalibrated Enable
Questionable Integrity Uncalibrated Event Query
Questionable Integrity Uncalibrated Negative Transition
Sequence Analyzer User's & Programmer's Reference
79