5 System Functions
System
Show Alignment Statistics
Shows alignment information you can use to ensure that the instrument is operating in a specific manner.
The Show Alignment Statistics screen is where you can view time and temperature information.
Values which are displayed are only updated when the Show Alignment Statistics screen is invoked, they
are not updated while the Show Alignment Statistics screen is being displayed. The remote commands
that access this information obtain current values.
An example of the Show Alignment Statistics screen would be similar to:
A successful Align Now, RF will set the Last Align RF temperature to the current temperature, and reset
the Last Align RF time. A successful Align Now, All or Align Now, All but RF will set the Last Align Now All
temperature to the current temperature, and reset the Last Align Now All time. A successful Align Now, All
will also reset the Last Align RF items if the RF portion of the Align Now succeeded.
Key Path
System, Alignments
Mode
All
Notes
The values displayed on the screen are only updated upon entry to the screen and not updated while
the screen is being displayed.
Initial S/W Revision
Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode
All
Remote Command
:SYSTem:PON:TIME?
Example
:SYST:PON:TIME?
Notes
Value is the time since the most recent start-up in seconds.