6 List Sequence Measurements
Meas Setup
Range
On | Off
Initial S/W Revision
A.05.00
Help Map ID
35278
Basic Discrete PAvT Setup
This measurement returns the amplitude, phase and frequency errors of each step. For the first step,
frequency errors are reference to the frequency set for the acquisition and phase are taken as zero. The
amplitude, frequency and phase errors of remained steps are reference to first step. Because Basic
Discrete PAvT needs to set the channel filter, it is exclusive with other measurements. It means that only
Basic Discrete PAvT is allowed for all analyses of this acquisition once Basic Discrete PAvT measurement is
selected for one analysis of this acquisition.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup
Mode
Sequence Analyzer
Initial S/W Revision
A.05.00
Help Map ID
35279
Interval Count
This allows you to specify the total number of steps to be measured.
parameter_
table_
23.52941
52.94118
Key Path
SCPI Only
Remote
Command
[:SENSe]:LSEQuencer:PCALibration::STEP:COUNt <integer>
[:SENSe] :LSEQuencer:PCALibration::STEP:COUNt?
Example
:LSEQ:PCAL:STEP:COUN 25
Notes
If the number is greater than the number of interval center or the number of the interval width, then any new
interval center and interval width will use the previous step interval center or interval width value.
Preset
1
State Saved
Saved in instrument state.
Range
1 to 512
Min
1
Max
512
Initial S/W
Revision
A.05.00
Help Map ID
0