6 List Sequence Measurements
List Sequence Measurements
11. Trigger set to output Int trig, but the test set’s source or analyzer is not set up to accept it.
12. Settings Conflict: Freq for acquisition > 3.6 GHz, hence the electronic attenuator cannot be used.
Validation Warnings/Advisories:
1. Warning: Meas Bitmap and Expected Power settings are in conflict. For example, EVM accuracy is in
question, due to the low power level. EVM accuracy is only guaranteed for power levels greater than –
65 dBm for RFIO paths and –70 dBm for RF In path.
2. Expected Power is in conflict with Instrument Gain setting.
3. Transition Time is not long enough for the Acquisition change.
If the file passes this test, errors may be generated at run time for the Sequence. In this case, the error
messages are sent to the error queue. The Sequence may or may not stop when an error is detected,
depending on whether you have selected “Abort on error”. Additionally, the error message contains the
Acquisition Number and Analysis Step Number (if applicable) where the error occurred.