6 List Sequence Measurements
List Sequence Measurements
parameter_table_23.52941
52.94118
Key Path
SCPI Only
Mode
Sequence Analyzer
Remote Command
[:SENSe]:LSEQuencer:ASETup:GSM:IEAVerage ON|OFF|1|0
[:SENSe]:LSEQuencer:ASETup:GSM:IEAVerage?
Example
LSEQ:ASET:GSM:IEAV 1
LSEQ:ASET:GSM:IEAV?
Notes
This setting is only in scope when the average number is greater than 1.
Preset
OFF
State Saved
Saved in instrument state.
Range
ON | OFF | 1 | 0
Initial S/W Revision
A.05.00
Help Map ID
0
PVT Time Offsets
The user may specify 12 offset time values, for which power levels can be queried. See
for details. The measurement offset results may be used by the user to perform a statistical analysis of the
burst shape, and is derived from the Agilent 8960 Series Wireless Communications Test Set.
The PVT measurement has a fixed number of 12 offset time points for the burst captured by the List
Sequencer. The offset times are referenced to the centre of symbol 0 for that burst (the start of the useful
part of the burst). The power levels relative to the useful part at these time offsets are calculated using
linear interpolation from the captured PVT IQ samples.
The Time Offsets are set for the entire Sequence.
parameter_
table_
23.52941
52.94118
Key Path
SCPI Only
Mode
Sequence Analyzer
Remote
Command
[:SENSe]:LSEQuencer:ASETup:GSM:PVTime:TIME[:OFFSet] <time>, <time>, <time>,
<time>, <time>, <time>, <time>, <time>, <time>, <time>, <time>, <time>
[:SENSe]:LSEQuencer:ASET:GSM:PVT:TIME[:OFFSet]?
Example
LSEQ:ASET:GSM:PVT:TIME –28us, –18us, –10us, 0, 2us, 4us, 538.2us, 540.2us, 542.8us, 552.8us, 560.8us,
570.8us
Preset
–2.8e–5, –1.8e–5, –1e–5, 0, 2e–6, 4e–6, 5.382e–4, 5.402e–4, 5.428e–4, 5.528e–4, 5.608e–4, 5.708e–4
State Saved
Saved in instrument state.
Range
–50 us to 593 us
Initial S/W
Revision
A.05.00
296
Sequence Analyzer User's & Programmer's Reference