6 List Sequence Measurements
Source
1000 steps, then insert more step will be rejected and popup error –221, “Setting Conflict; Cannot insert
more steps, maximum number of steps reached”
Key Path
Source, List Sequencer, List Sequencer Setup
Notes
No remote command, front panel only. If the list already contains the maximum limit of 1000 steps,
no operation will be made after pressing this key.
Initial S/W Revision
A.05.00
Delete Step
Allows you to delete the current step. Deleting a step will automatically decrease the Step Count
parameter by 1. If sequence only has one step left, delete step will be rejected and popup error –221,
“Setting conflict; Cannot delete current step, minimum number of steps reached”
Key Path
Source, List Sequencer, List Sequencer Setup
Notes
No remote command, Front Panel key only.
If the list already contains the minimum limit of 1 step, no operation will be made after pressing this
key
Initial S/W Revision
A.05.00
Clear List
Allows you to clear the list. Clearing the list sets the number of steps to the default value of 1 and sets the
parameters for the only step to their default values.
Key Path
Source, List Sequencer, List Sequencer Setup
Initial S/W Revision
A.05.00
Step Trigger
Allows access to the sub-menu for selecting the trigger input for the current step.
Key Path
Source, List Sequencer, List Sequencer Setup
Remote Command
:SOURce:LIST:STEP[1]|2|3…1000:SETup:INPut:TRIGger IMMediate | INTernal
| EXTernal2 | KEY | BUS | EXTernal4
:SOURce:LIST:STEP[1]|2|3…1000:SETup:INPut:TRIGger?
Example
:SOUR:LIST:STEP2:SET:INP:TRIG BUS
:SOUR:LIST:STEP2:SET:INP:TRIG?
Notes
SCPI is supported after A.09.40
Preset
Free Run
Range
Free Run | Internal | Manual (Trigger Key) | Bus | External 2| EXTernal4
Initial S/W Revision
A.05.00
Sequence Analyzer User's & Programmer's Reference