6 List Sequence Measurements
Source
Source
Opens a menu of keys that access various source configuration menus and settings. In the test set,
pressing this key also causes the central view area to change and display the Source Control Main view.
Key Path
Front-panel key
RF Output
This parameter sets the source RF power output state.
Key Path
Source
Remote Command
:OUTPut[:EXTernal][:STATe] ON|OFF|1|0
:OUTPut[:EXTernal][:STATe]?
Example
OUTP OFF
OUTP?
Notes
The EXTernal node is shown in RD text so the SCPI remains the same between internal and external
source control. However, for EXT we do not wish to document this node to the customer since we
are controlling the internal source rather than the external source.
This setting is for the independent mode and has no effect on the
. If
the
is set to ON, the list sequencer controls the source output and this key
will be grayed-out. And this setting will be none-forceful grey out on front panel to indicate out-of-
scope. Non-forceful means user still can change this setting by SCPI but cannot change on front
panel. When set to OFF will make source leave list sequencer and this setting will be black out and
take effect immediately.
When the RF Output is ON, an “RF” annunciator is displayed in the system settings panel. When the
RF Output is turned Off, the RF annunciator is cleared. If the
is set to ON,
the “RF” annunciator will be replaced by “SEQ” in the system settings panel, indicating that the
output is controlled by the list sequencer.
Preset
Off
Range
On | Off
Initial S/W Revision
A.05.00
Amplitude
Allows you to access the Amplitude sub-menu.
Key Path
Source
Notes
The sub-menu under this button is for independent mode and has no effect on
. If the
is set to ON, the list sequencer controls the source output
and this key will be grayed-out on front panel to indicate out-of-scope. When you set
to Off will make source leave list sequencer and this button will be black out.
Initial S/W Revision
A.05.00
Sequence Analyzer User's & Programmer's Reference