2 Programming the Test Set
STATus Subsystem
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature:ENABle <integer>
:STATus:QUEStionable:TEMPerature:ENABle?
Example
STAT:QUES:TEMP:ENAB 1 Reference Oscillator Oven Cold will be reported to the Temperature
Summary of the Status Questionable register.
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Temperature Event Query
This query returns the decimal value of the sum of the bits in the Questionable Temperature Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature[:EVENt]?
Example
STAT:QUES:TEMP?
Preset
0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Temperature Negative Transition
This command determines which bits in the Questionable Temperature Condition register will set the
corresponding bit in the Questionable Temperature Event register when the condition register bit has a
negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that you want
to enable.
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature:NTRansition <integer>
:STATus:QUEStionable:TEMPerature:NTRansition?
Example
STAT:QUES:TEMP:NTR 1 Reference Oscillator Oven not cold will be reported to the Temperature
Summary of the Status Questionable register.
Preset
0
Sequence Analyzer User's & Programmer's Reference