F-20
Contact Check Function
2400 Series SourceMeter
®
User’s Manual
TOUT <NRf>
:TRIGger:SEQuence2:TOUT <NRf>
Specify contact check timeout
Parameters
<NRf> = 0.00000 to 999.99990(sec) Contact check timeout
Query
:TOUT?
Query contact check timeout
Description
This command programs the contact check timeout interval. When con-
tact event detection is enabled (
“SOURce <name>,” page F-19
), the
SourceMeter will wait the specified timeout period before issuing an
open lead failure and moving on to the next measurement. If the timeout
interval is set to 0 with contact event detection enabled, the unit will
wait indefinitely.
Contact check defaults
Table F-5
summarizes default settings for various contact check modes along with the
equivalent remote commands.
Table F-5
Contact check defaults
Mode
Default setting*
Remote command*
Contact check mode
Contact check threshold resistance
Limit 4 test
Limit 4 Digital I/O failure pattern
Contact check event detection
Contact check timeout
Disabled
50
Ω
Disabled
15
Disabled
0
:SYSTem:CCHeck OFF
:SYStem:CCHeck:RESistance 50
:CALCulate2:LIMit4:STATe OFF
:CALCulate2:LIMit4:SOURce2 15
:TRIGger:SEQuence2:SOURce IMMediate
:TRIGger:SEQuence2:TOUT 0
*Bench (:SYSTem:PRESet) and GPIB (*RST) defaults are the same.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...