12-6
Limit Testing
2400 Series SourceMeter
®
User’s Manual
Figure 12-2
Grading mode limit testing
Perform
Limit 1 Test
?
Yes
No
Pass
?
No
Yes
Display
“FAIL”
Perform
Limit 2 Test
?
Yes
No
Pass
?
No
Yes
Perform
Limit 3, 5-12
Tests
?
Yes
No
Pass
?
No
Yes
Another
Test Cycle
?
No
Yes
Start
Turn Output ON and press LIMIT key.
Output First
Fail Pattern
Test
Another
Device
?
No
Yes
Stop
Press LIMIT
Yes
No
Wait for SOT pulse
from handler
Perform Source-
Measure action
Display
“FAIL”
Display
“FAIL”
Binning
Control
Binning
Control
Binning
Control
First
Failure
?
First
Failure
?
First
Failure
?
End
End
End
Yes
Yes
Yes
Output Limit 1
Fail Pattern
Immediate
Output Limit 2
Fail Pattern
Immediate
Output Limit 3
Fail Pattern
Immediate
Store Limit 1 Fail
Pattern in Memory
Store Limit 2 Fail
Pattern in Memory
Store Limit 3, 5-12 Fail
Pattern in Memory
No
No
No
Any
Failures
?
Display
“PASS”
Yes
Binning
Control
End
Any
Failures
?
Immediate
Output Pass
Pattern
Note:
Limit 4 (contact option) not shown. See Appendix F.
No
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...