3-6
Basic Source-Measure Operation
2400 Series SourceMeter
®
User’s Manual
Compliance limit
When sourcing voltage, the SourceMeter can be set to limit current. Conversely, when
sourcing current, the SourceMeter can be set to limit voltage. The SourceMeter output will
not exceed the compliance limit.
Table 3-2
summarizes compliance limits according to
range. See
Section 6
for more details on compliance limit.
Table 3-2
Compliance limits
2400
2410
2420
Measure
range
Maximum
compliance
value
Measure
range
Maximum
compliance
value
Measure
range
Maximum
compliance
value
200mV
2V
20V
200V
±210mV
±2.1V
±21V
±210V
200mV
2V
20V
1000V
±210mV
±2.1V
±21V
±1.1kV
200mV
2V
20V
60V
±210mV
±2.1V
±21V
±63V
1µA
10µA
100µA
1mA
10mA
100mA
1A
±1.05µA
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
1µA
10µA
100µA
1mA
20mA
100mA
1A
±1.05µA
±10.5µA
±105µA
±1.05mA
±21mA
±105mA
±1.05A
10µA
100µA
1mA
10mA
100mA
1A
3A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±3.15A
2425/2430
2440
Measure
range
Maximum
compliance
value
Measure
range
Maximum
compliance
value
200mV
2V
20V
100V
±210mV
±2.1V
±21V
±105V
200mV
2V
10V
40V
±210mV
±2.1V
±10.5V
±42V
10µA
100µA
1mA
10mA
100mA
1A
3A/10A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
*
10µA
100µA
1mA
10mA
100mA
1A
5A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±5.25A
*
±3.15A (2425 and 2430 DC mode)
±10.5A (Pulse mode; 2430 only)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...