2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference
18-15
[:SOURce[1]]
:VOLTage
[:LEVel]
Set V-Source level (in volts):
✓
[:IMMediate]
Set specified level immediately:
✓
[:AMPLitude] <n>
Specify voltage level.
1
0
✓
✓
3
[:AMPLitude]?
Query voltage level.
✓
:TRIGgered
Set specified level when triggered:
✓
[:AMPLitude] <n>
Specify voltage level.
1
0
✓
[:AMPLitude]?
Query voltage level.
✓
:SFACtor <n>
Set voltage scaling factor
(-999.9999e+18 to +999.9999e+18).
0
✓
3
:STATe <b>
Enable/disable voltage scaling factor. OFF
✓
3
:STATe?
Query voltage scaling factor state.
:SFACtor?
Query voltage scaling factor.
:PROTection
Path to limit output voltage:
✓
[:LEVel] <NRf>
Specify voltage limit level.
1
(Note 2)
✓
[:LEVel]?
Query voltage limit.
✓
:TRIPped?
Voltage limit detected: 1 (yes), 0 (no).
✓
:STARt <n>
Specify start level for V-sweep.
1
0
✓
:STARt?
Query start level for voltage sweep.
✓
:STOP <n>
Specify stop level for V-sweep.
1
0
✓
:STOP?
Query stop level for voltage sweep.
✓
:STEP <n>
Specify step value for V-sweep.
1
0
:STEP?
Query step value for voltage sweep.
:SPAN <n>
Specify span.
1
0
✓
:SPAN?
Query span.
✓
:CENTer <n>
Specify center point.
1
0
✓
:CENTer?
Query center point.
✓
:SOAK <NRf>
Set first sweep point soak time (0.00000 to
9999.999s).
0.00000
✓
:SOAK?
Query soak time.
✓
1. The parameter values for this command are provided with the detailed command reference information. Detailed information on
SCPI commands follows the last command summary table in this section.
2. SYSTem:PRESet default is 40, *RST default is NONE.
3. If source V is active.
Table 18-7 (continued)
SOURce command summary
Command
Description
Default
parameter SCPI
Source
memory
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...