18-98
SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
Now add the LogStep value to Log
10
(Start) and to each subsequent result. This will create
a list of Log
10
Values. Next take the anti-log of each Log
10
Value to get the actual sweep
values:
Value#
Log
10
Value
Sweep Value
1
-3.000000
0.001000
2
-2.872033
0.001343
3
-2.744067
0.001803
4
-2.616100
0.002420
5
-2.488134
0.003250
6
-2.360167
0.004363
7
-2.232201
0.005859
8
-2.104234
0.007866
9
-1.976268
0.010562
10
-1.848301
0.014181
11
-1.720335
0.019040
12
-1.592368
0.025564
13
-1.464402
0.034324
14
-1.336435
0.046086
15
-1.208469
0.061877
16
-1.080502
0.083080
17
-0.952536
0.111549
18
-0.824569
0.149772
19
-0.696603
0.201093
20
-0.568636
0.270000
Current list
The Previous Log Current Sweep can instead be performed by using the sweep values in a
Current List as follows:
*RST
SOUR:FUNC:MODE CURR
SOUR:LIST:CURR 0.001,0.001343,0.001803,0.002420,0.003250,0.004363,0.005859
SOUR:LIST:CURR:APP 0.007866,0.010562,0.01418,0.019040,0.025564,0.03424
SOUR:LIST:CURR:APP
0.046086,0.061877,0.083080,0.111549,0.149772,0.201093,0.27
SOUR:LIST:CURR:POIN? (returns 20)
TRIG:COUN 20
SOUR:CURR:MODE LIST
OUTP ON
INIT
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...