![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086568.webp)
2400 Series SourceMeter
®
User’s Manual
Contact Check Function
F-15
Command
Description
:FORMat:ELEMents:CALCulate STATus
Bit 18 (262144) of the status word returned by READ?,
FETCh?, MEASure?, MEASure:VOLTage?, MEA-
Sure:CURRent?, MEASure:RESistance?, TRACe:DATA?,
CALCulatel :DATA?, or :CALCulate2:DATA? signifies the
status of contact check (1 = open lead, 0 = OK). Also, bit 8 set
if LIMit4 is enabled.
:SYSTem:RSENse ON
Enable 4-wire
remote sense for contact check.
:SYSTem:RSENse?
Query sense selection.
:SYSTem:CCHeck ON |
OFF
Enable/disable contact checking.
:SYSTem:CCHeck?
Query state of contact check test.
:SYSTem:CCHEck:RESistance <NRf>
Select threshold resistance (0 to 60):
<NRf> = 0 to 2 Selects 2
Ω
threshold resistance
= >2 to
15
Selects
15
Ω
threshold resistance
= >15 to 60 Selects 50
Ω
threshold resistance
:SYSTem:CCHeck:RESistance?
Query threshold resistance.
:TRIGger:SEQuence2:SOURce <name>
Enable or disable contact check event detection; <name> =
CCHeck (enable) or
IMMediate
(disable).
:TRIGger:SEQuence2:SOURce?
Query state of event detection.
:TRIGger:SEQuence2:TOUT <NRf>
Specify contact check timeout in seconds (
0.00000
to
999.99990).
:TRIGger:SEQuence2:TOUT?
Query contact check timeout setting.
*RST and SYSTem:PRESet parameter defaults are shown in boldface.
Table F-3 (continued)
Contact check remote commands
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...