12-14
Limit Testing
2400 Series SourceMeter
®
User’s Manual
Basic binning systems
Two basic binning systems are shown in
Figure 12-7
and
Figure 12-8
. Both systems
require a handler to physically place the device packages in the appropriate bins. The han-
dler is controlled by the SourceMeter via the Digital I/O port.
Single-element device binning
Figure 12-7
shows a basic binning system for single-element devices (i.e., resistors). After
all programmed testing on the DUT is completed, the pass/fail digital output information
is sent to the component handler, which then places the DUT in the appropriate bin. The
component handler selects the next DUT, and the testing process is repeated.
Figure 12-7
Binning system single-element devices
Handler
Dig
In
DUT
HI
LO
IN/OUT
SourceMeter
Dig
I/O