2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference
18-69
2425 and 2430 DC Mode
<n> =
-3.15 to 3.15
Current compliance limit
-105 to 105
Voltage compliance limit
DEFault
105uA, 21V
MINimum
-3.15A, -105V
MAXimum
3.15A, 105V
2430 Pulse Mode
<n> =
-10.5 to 10.5
Current compliance limit
-105 to 105
Voltage compliance limit
DEFault
105uA, 21V
MINimum
-10.5A, -105V
MAXimum
10.5A, 105V
2440
<n> =
-5.25 to 5.25
Current compliance limit
-42 to 42
Voltage compliance limit
DEFault
105uA, 21V
MINimum
-5.25A, -42V
MAXimum
5.25A, 42V
Query
:LEVel?
Query compliance value
:LEVel? DEFault
Query *RST default compliance
:LEVel? MINimum
Query minimum allowable compliance
:LEVel? MAXimum
Query maximum allowable compliance
Description
This command is used to set compliance limits. A current compliance
limit is set for the V-Source, and a voltage compliance limit is set for the
I-Source. The SourceMeter cannot source levels that exceed these speci-
fied limits.
The :SENSe:CURRent:PROTection[:LIMit] command is used to set the
current compliance for the V-Source and the :SENSe:VOLTage:
PROTection[:LIMit] command is used to set the voltage compliance for
the I-Source.
NOTE
You cannot set compliance less than 0.1% of the present measurement range.
RSYNhronize <b>
[:SENSe[1]]:CURRent[:DC]:PROTection:RSYNchronize <b>
Control current measure and compliance
range sync
[:SENSe[1]]:VOLTage[:DC]:PROTection:RSYNchronize <b>
Control voltage measure and compliance
range sync
Parameters
<b> =
OFF
Disable range sync
ON
Enable range sync
Description
Turning this feature on will make the measurement range follow the
compliance range setting when autoranging is off. When the compliance
value is set, the measurement range will be on the same range as the
compliance setting. The two commands above change one setting;
changing the state of one will automatically change the other.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...