![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086043.webp)
1-6
Getting Started
2400 Series SourceMeter
®
User’s Manual
Model 2420:
•
Source voltage from 5µV to 63V; measure voltage from 1µV to 63.3V.
•
Source current from 500pA to 3.15A; measure current from 100pA to 3.165A.
•
Measure resistance from 10µ
Ω
(<10µ
Ω
in manual ohms) to 21.1M
Ω
.
•
Maximum source power is 66W.
Models 2425 and 2430:
•
Source DC or pulse voltage from 5µV to 105V; measure voltage from 1µV to
105.5V.
•
Source DC current from 500pA to 3.15A; measure DC current from 100pA to
3.165A.
•
Source pulse current from 500pA to 10.5A; measure pulse current from 100pA to
10.55A. (Model 2430 only.)
•
Measure resistance from 10µ
Ω
(<10µ
Ω
in manual ohms) to 21.1M
Ω
.
•
Maximum DC source power is 110W.
•
Maximum pulse source power is 1.1kW. (Model 2430 only.)
Model 2440:
•
Source voltage from 5µV to 42V; measure voltage from 1µV to 42V.
•
Source current from 500pA to 5.25A; measure current from 100pA to 5.25A.
•
Measure resistance from 10µ
Ω
(<10µ
Ω
in manual ohms) to 21.1M
Ω
.
•
Maximum source power is 66W.
NOTES
The Models 2400, 2410, 2420, 2425, 2430, and 2440 are Y2K compliant.
Models with a “-C” suffix have the Contact Check option. See
Appendix F
.
Some additional capabilities of the SourceMeter include:
•
Concurrent measurements of all three functions over the remote interface.
•
Source-measure sweep capabilities (linear and logarithmic staircase sweeps,
source sweep list of up to 2500 points, memory sweep of up to 100 instrument set-
ups).
•
6-wire
Ω
measurement with programmable I-source or V-source with V or I clamp.
•
4-quadrant source and sink operation.
•
Up to 12 stages of limit testing with a built-in comparator for pass/fail testing.
•
Digital I/O for stand-alone binning operations or interface to component handler.
•
Programming language and remote interfaces — The SourceMeter uses the SCPI
programming language and two remote interface ports (IEEE-488/GPIB and
RS-232C).
•
Trigger-Link interface to Keithley Series 7000 switching hardware.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...