5-14
Pulse Mode Operation (Model 2430 only)
2400 Series SourceMeter
®
User’s Manual
Step 7: Turn output on.
Turn the output on by pressing the TRIG key or the ON/OFF OUTPUT key. The OUTPUT
indicator will be on whenever a pulse is being sourced. During each pulse off-time, the
output will be off (indicator off).
If configured to output a finite number of pulses, the output will disable after the last pulse
is sourced. While sourcing pulses, you can disable the output at any time by pressing the
ON/OFF OUTPUT key.
Pulse-ohms measurements
NOTE
Ohms source readback cannot be used for the Pulse Mode. Make sure that the
source is not in compliance when taking pulse-ohms readings.
Section 4
covers
“
Ohms source readback
.”
Auto and Manual pulse-ohms measurements
While in the Pulse Mode, the Model 2430 can perform auto or manual ohms measure-
ments. After selecting and configuring the Pulse Mode as explained in this section, per-
form the “
Auto ohms measurements
” procedure or the “
Manual ohms measurements
”
procedure in
Section 4
.
For remote operation, see
Section 4
, “
Remote ohms programming
,” for the commands for
basic ohms measurements.
Offset-compensated pulse-ohms measurements
While in the Pulse Mode, offset-compensated ohms is available as a math function
(FCTN). After selecting and configuring the Pulse Mode as explained in this section, refer
to
Section 8
, “
Math operations
,” to perform offset-compensated ohms measurements.
For remote operation, see
Section 8
, “
Remote math operations
,” for the commands to per-
form offset-compensated ohms.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...