2400 Series SourceMeter
®
User’s Manual
Contact Check Function
F-17
Contact check command reference
Configure and control contact check
CCHeck <b>
:SYSTem:CCHeck <b>
Enable or disable contact check.
Parameters
<b> =
0 or OFF
Disable contact check
1 or ON
Enable contact check
Query
:CCHeck?
Query state of contact check
Description
This command is used to enable or disable contact check. When using
contact check, remote sensing must be enabled (:SYST:RSEN ON).
RESistance <NRf>
:SYSTem:CCHeck:RESistance <NRf>
Set contact check threshold resistance.
Parameters
<NRf> = 0 - 60
2
Ω,
15
Ω
, or 50
Ω
threshold resistance
0 to 2 = 2
Ω
>2 to 15 = 15
Ω
>15 to 60 = 50
Ω
Query
:RESistance?
Query contact check threshold resistance
Description
This command sets the contact threshold resistance value. (Factory
default = 50
Ω
.) The contact check function will indicate a failure when
the DUT resistance is greater than the programmed value.
Configure and control contact check limit test
STATe <b>
:CALCulate2:LIMit4:STATe <b>
Control LIMIT 4 test
Parameters
<b> =
1 or ON
Enable Limit 4 test
0 or OFF
Disable Limit 4 test
Query
:STATe?
Query state of Limit 4 test
Description
This command enables or disables the LIMIT 4 (contact check) test.
When the limit test is enabled, the Digital I/O port bit pattern can be
controlled by the results of the test. (See
“SOURce2 <NRf> |<NDN>,”
page F-18
.)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com