4-4
Ohms Measurements
2400 Series SourceMeter
®
User’s Manual
Selecting ohms measurement method
On power-up, auto ohms is the default method for the ohms function. Perform the follow-
ing steps to check and/or change the ohms measurement method:
1.
Press CONFIG and then
Ω
to display the ohms configuration menu.
2.
Using the EDIT keys, place the cursor (flashing menu item) on SOURCE and press
ENTER.
NOTE
Cursor position indicates the presently selected ohms measurement method. To
retain this selection, use the EXIT key to back out of the menu structure and skip
the next two steps.
3.
To change the measurement method, place the cursor on the alternate selection
(AUTO or MANUAL), and press ENTER.
4.
Press EXIT to exit from the menu structure.
Auto ohms measurements
Perform the following steps to perform auto ohms measurements.
NOTE
The following procedure assumes that the SourceMeter is already connected to
the DUT as explained in
Section 2
, “
Connections
.”
WARNING
To prevent electric shock, do not make or break connections to the
SourceMeter with the output on. If on, press the ON/OFF OUTPUT
key to turn the output off.
1.
Select ohms measurement function.
Press MEAS
Ω
to select the ohms measurement function.
2.
Select auto ohms measurement method.
Select the auto ohms measurement method as follows:
•
Press CONFIG then
Ω.
•
Select SOURCE, then press ENTER.
•
Select AUTO, then press ENTER.
•
Press EXIT to return to normal display.
Note that the SourceMeter will be configured to Source I and Measure V. Also, the
I-Source level and V-Compliance limit are based on the measurement range and
cannot be edited.
NOTE
Use the manual ohms mode and the V-source method when high-speed settling is
required.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...