Pulse Mode Operation (Model 2430 only)
2400 Series SourceMeter
®
User’s Manual
Pulse Mode configuration
Front panel Pulse Mode configuration
Select Pulse Mode, and set pulse width and pulse delay
1.
Press CONFIG then SOURCE V or I.
2.
Select SHAPE from the displayed choices, then press ENTER.
3.
Select PULSE from the displayed choices, then press ENTER. Selecting the Pulse
Mode enables the pulse width and pulse delay items of the menu.
4.
Select PULSE WIDTH from the displayed choices, then press ENTER.
5.
Enter the desired pulsewidth value (0.15ms to 5.00ms), then press ENTER. Note
that the pulse width internally will limit to 2.5ms on the 10A range (source or
measure).
6.
Select DELAY from the displayed choices, then press ENTER.
7.
Enter the desired pulse delay value (0 to 9999.99872 sec), then press ENTER.
8.
Press EXIT to return to the normal display.
Set pulse measurement speed
Once the Pulse Mode is selected, the available NPLC values become accessible to set
pulse measurement speed.
1.
Press SPEED to display the pulse speed menu.
2.
Select the speed setting (0.004 to 0.100) and press ENTER.
Set pulse count
The arm count and trigger count determine how many pulses will be sourced. In general,
the number of pulses to be sourced is the product of the arm count and trigger count. For
example, if the arm count is two and the trigger count is five, ten pulses will be sourced
when the output is turned on. However, with the arm count set to infinite, the Model 2430
will continuously source pulses when the output is turned on. Note that the product of
finite value counts cannot exceed 2500. See
Section 11
, “
Trigger models
,” for details.
Perform the following steps to set the arm count:
1.
Press CONFIG and then TRIG to display the trigger configuration menu.
2.
Select ARM LAYER from the displayed choices and press ENTER.
3.
Select COUNT from the displayed choices and press ENTER.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...