4-2
Ohms Measurements
2400 Series SourceMeter
®
User’s Manual
Ohms configuration menu
NOTE
For the Model 2430 Pulse Mode, offset-compensated ohms cannot be enabled
from the ohms configuration menu. However, offset-compensated ohms is avail-
able as a math function (
Section 8
, “
Math operations
”).
To access the ohms configuration menu, press CONFIG then OHMS. Use
Section 1
,
“
Rules to navigate menus
,” to select the various items in the menu tree, which is shown in
Figure 4-1
. Menu items include:
•
SOURCE — Select AUTO or MANUAL source mode.
•
SENSE MODE — Select 2-WIRE or 4-WIRE sense mode.
•
GUARD — Choose OHMS or CABLE guard.
•
OFFSET COMPENSATION — Enable or disable offset-compensated ohms (not
valid for Model 2430 Pulse Mode).
•
SRC RDBK — Enable or disable source readback mode.
The following paragraphs discuss each of these aspects in detail.
Figure 4-1
Ohms configuration menu tree
CONFIG
Ω
GUARD
SOURCE
SENSE
MODE
OFFSET
COMPENSATION
SRC
RDBK
MANUAL
AUTO
4-WIRE
2-WIRE
CABLE
OHMS
DISABLE
ENABLE
DISABLE
ENABLE