2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference
18-11
[:SENSe[1]]
:CURRent[:DC]
Path to configure current:
✓
:RANGe
Configure measurement range:
✓
[:UPPer]
<n>|UP|DOWN
Select range by specifying the expected
current reading.
1
1.05e-4
✓
✓
4
[:UPPer]?
Query range.
✓
:AUTO <b>
Enable or disable auto range.
2
ON
✓
✓
4
:AUTO?
Query auto range.
✓
:ULIMit?
Returns I compliance range.
✓
:LLIMit <n>
Set auto ranging lower limit (-105e-6
to 105e-6).
1e-6
✓
:LLIMit?
Query auto ranging lower limit.
✓
:NPLCycles <n>
Specify integration rate (in line cycles):
0.01 to 10.
3
1.0
✓
✓
4
:NPLCycles?
Query integration rate.
✓
:PROTection
Path to configure current compliance:
✓
[:LEVel] <n>
Specify current limit for V-Source.
1
1.05e-4
✓
✓
4
[:LEVel]?
Query current compliance limit.
✓
:TRIPped?
In current compliance: 1 (yes), 0 (no).
✓
:RSYNchronize <b>
Enable or disable measure and
compliance range synchronization.
OFF
:RSYNchronize?
Query range synchronization.
:VOLTage[:DC]
Path to configure volts:
✓
:RANGe
Configure measurement range:
✓
[:UPPer]
<n>|UP|DOWN
Select range by specifying the expected
voltage reading.
1
21
✓
✓
5
[:UPPer]?
Query range.
✓
:AUTO <b>
Enable or disable auto range.
2
ON
✓
✓
5
:AUTO?
Query auto range.
✓
:ULIMit?
Returns V compliance range.
✓
:LLIMit <n>
Set auto range lower limit.
1
0.21
✓
:LLIMit?
Query auto range lower limit.
✓
:NPLCycles <n>
Specify integration rate (in line cycles):
0.01 to 10.
3
1.0
✓
✓
5
:NPLCycles?
Query integration rate.
✓
:PROTection
Path to configure voltage compliance:
✓
[:LEVel] <n>
Specify voltage limit for I-Source.
1
21
✓
✓
5
1. The parameter values for this command are provided with the detailed command reference information. Detailed information on SCPI commands fol-
lows the last command summary table in this section.
2. For the Model 2430 Pulse Mode, auto range is invalid (error +831).
3. For the Model 2430 Pulse Mode, <n> = 0.004 to 0.100.
4. If source V is active.
5. If source I is active and auto ohms is disabled.
Table 18-6 (continued)
SENSe command summary
Command
Description
Default
parameter SCPI
Source
memory
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...