18-112 SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
Error queue
NOTE
See
Section 15
for details on the error queue.
[:NEXT]?
:SYSTem:ERRor[:NEXT]?
Read oldest error (code and message)
Description
As error and status messages occur, they are placed in the Error Queue.
The Error Queue is a first-in, first-out (FIFO) register that can hold up
to 10 messages. After sending this command and addressing the
SourceMeter to talk, the oldest message is sent to the computer and is
then removed from the queue.
NOTE
The :STATus:Queue? command performs the same function as
:SYSTem:ERRor[:NEXT]?. (See
“STATus subsystem,” page 18-104
.)
ALL?
:SYSTem:ERRor:ALL?
Read all errors (codes and messages)
Description
This query command is similar to the [:NEXT]? command except that
all messages in the Error Queue are sent to the computer when the
SourceMeter is addressed to talk. All messages are removed from the
queue.
COUNt?
:SYSTem:ERRor:COUNt?
Return the number of errors
Description
After sending this command and addressing the SourceMeter to talk, a
decimal number will be sent to the computer. That is the number of mes-
sages in the Error Queue.
CODE[:NEXT]?
:SYSTem:ERRor:CODE[:NEXT]?
Read oldest error (code only)
Description
This command is identical to the [:NEXT]? command, except only the
code is returned. The message itself is not returned. The error is cleared
from the queue.
CODE:ALL?
SYSTem:ERRor:CODE:ALL?
Read all errors (codes only)
Description
This query command is identical to the :ALL? command, except only
the codes are returned. The actual messages are not returned. All errors
are cleared from the queue.
CLEar
:SYSTem:CLEar
Clear Error Queue
Description
This action command is used to clear the Error Queue of messages.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...