F-12
Contact Check Function
2400 Series SourceMeter
®
User’s Manual
Using contact check
1.
Make connections to the DUT (
Figure F-3
).
2.
Press the MENU key, select A/D CTRL, then choose CONT CHK, and press
ENTER.
3.
Select ENABLE, then press ENTER.
4.
At the CONT-CHK RESISTANCE prompt, select the desired contact check thresh-
old resistance (2
Ω
, 15
Ω
, or 50
Ω
), then press ENTER. (See
Table F-1
for
recommendations.)
5.
Use EXIT to back out of the menu structure.
6.
Press CONFIG then
Ω
, then select SENSE MODE, and press ENTER.
Menu item*
Description
MENU
Main menu.
MAIN MENU
A/D CTRL
A/D control selection.
AUTO ZERO
LINE FREQ
NPLC CACHE
Control NPLC caching.
CONT CHK
Enable/disable contact check.
DISABLE
Disable contact check.
ENABLE
Enable contact check, select 2
Ω
, 15
Ω
, or 50
Ω
resistance.
CONFIG TRIG
Configure triggering.
CONFIGURE TRIGGER
ARM LAYER
Configure arm layer.
TRIG LAYER
Configure trigger layer.
HALT
Return to idle.
CONTACT CHECK
Configure contact check event detection.
DISABLE
Disable contact check event detection.
ENABLE
Enable contact check event detection, program contact
check timeout (
0.00000
to 999.99990).
CONFIG SOURCE V or CONFIG MEAS
Ω
Configure 4-wire sense mode for contact check.
SENSE MODE
Select sense mode.
2 WIRE
2-wire sensing.
4 WIRE
4-wire sensing (use for contact check).
*See
Section 1
for complete menu listings, and
Section 11
and
Section 12
for triggering and limit tests. Factory defaults shown in
boldface.
Table F-2 (continued)
Contact check menu selections
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...