2400 Series SourceMeter
®
User’s Manual
Remote Operations
14-13
Angle brackets
< > — Angle brackets (< >) are used to denote a parameter type. Do not
include the brackets in the program message. For example:
:OUTPut <b>
The <b> indicates a Boolean-type parameter is required. Therefore, to
enable the selected source, you must send the command with the ON or
1 parameter as follows:
:OUTPut ON
:OUTPut 1
Query commands
This type of command requests (queries) the presently programmed status. It is identified
by the question mark (?) at the end of the fundamental form of the command. Most com-
mands have a query form:
:ARM:TIMer?
Queries the timer interval.
Most commands that require a numeric parameter (<n>) can also use the DEFault,
MINimum, and MAXimum parameters for the query form. These query forms are used to
determine the *RST default value and the upper and lower limits for the fundamental com-
mand. Examples are:
:ARM:TIMer? DEFault
Queries the *RST default value.
:ARM:TIMer? MINimum
Queries the lowest allowable value.
:ARM:TIMer? MAXimum
Queries the largest allowable value.
Case sensitivity
Common commands and SCPI commands are not case sensitive. You can use upper or
lower case and any case combination. Examples:
*RST
= *rst
:DATA?
= :data?
:SYSTem:PRESet =
:system:preset
NOTE
Using all upper case will result in slightly faster command response times.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...