SOUR:VOLT:STAR 1.0
SOUR:VOLT:STOP 10.0
SOUR:VOLT:STEP 1.0
SOUR:SWE:POIN? (returns 10)
TRIG:COUN 10
SOUR:VOLT:MODE SWE
OUTP ON
INIT
Voltage list
The previous Linear Voltage Sweep can instead be performed using a Voltage List as
follows:
*RST
SOUR:FUNC:MODE VOLT
SOUR:LIST:VOLT 1,2,3,4,5,6,7,8,9,10
SOUR:LIST:VOLT:POIN? (returns 10)
TRIG:COUN 10
SOUR:VOLT:MODE LIST
OUTP ON
INIT
Logarithmic current sweep
Logarithmic current sweep from 1mA to 270mA in 20 points:
*RST
SOUR:FUNC:MODE CURR
SOUR:SWE:SPAC LOG
SOUR:CURR:STAR .001
SOUR:CURR:STOP .27
SOUR:SWE:POIN 20
TRIG:COUN 20
SOUR:CURR:MODE SWE
OUTP ON
INIT
To determine the source values that will be generated:
Start:
0.001
Log
10
(Start):
-3
Stop:
0.270
Log
10
(Stop):
-0.568636
LogStep = (Log
10
(Stop) - Log
10
(Start)) / (SWE:POIN -1)
= (-0.568636 - (-3) / (20 - 1)
= 2.431364 / 19
= 0.127967
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com