18-78
SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
2420
<n> =
-3.15 to 3.15
Set I-Source amplitude (amps)
-63 to 63
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-3.15A or -63V
MAXimum
+3.15A or +63V
2425 and 2430 DC Mode
<n> =
-3.15 to 3.15
Set I-Source amplitude (amps)
-105 to 105
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-3.15A or -105V
MAXimum
+3.15A or +105V
2430 Pulse Mode
<n> =
-10.5 to 10.5
Set I-Source amplitude (amps)
-105 to 105
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-10.5A or -105V
MAXimum
+10.5A or +105V
2440
<n> =
-5.25 to 5.25
Set I-Source amplitude (amps)
-42 to 42
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-5.25A or -42V
MAXimum
+5.25A or +42V
Query
:CURRent?
Query programmed amplitude of I-Source
:CURRent? DEFault
Query *RST default amplitude
:CURRent? MINimum
Query lowest allowable amplitude
:CURRent? MAXimum
Query highest allowable amplitude
:VOLTage?
Query programmed amplitude of V-Source
:VOLTage? DEFault
Query *RST default amplitude
:VOLTage? MINimum
Query lowest allowable amplitude
:VOLTage? MAXimum
Query highest allowable amplitude
Description
This command is used to immediately update the amplitude of a fixed
source. This command is not valid if the list or sweep mode is selected.
NOTE
The sourcing :MODE command is used to select a fixed source. (See
“Select
sourcing mode,” page 18-74
.)
If a manual source range is presently selected, then the specified ampli-
tude cannot exceed that range. For example, if the V-Source is on the 2V
range (auto range disabled), you will not be able to set the V-Source
amplitude to 3V. In auto range, the amplitude can be set to any level that
is within the capabilities of the source.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...