2400 Series SourceMeter
®
User’s Manual Digital I/O Port, Safety Interlock, and Output Configuration
13-5
Remote digital output control
Use the :SOURce2:TTL <NRf> command to control the digital output line logic levels,
where <NRf> is the decimal value shown in
Table 13-1
. For example, send the following
command to set the output lines to L, H, L, H:
:SOUR2:TTL 5
Safety interlock
The digital I/O port provides an interlock line for use with a test fixture interlock switch.
When properly used, the OUTPUT of the SourceMeter will turn OFF when the lid of the
test fixture is opened. See
Section 2
, “
Connections
,” for important safety information
when using the test fixture interlock.
Table 13-1
Digital output line settings
OUT 4
OUT 3
OUT 2
OUT 1
Decimal
value*
L
L
L
L
L
L
L
L
H
H
H
H
H
H
H
H
L
L
L
L
H
H
H
H
L
L
L
L
H
H
H
H
L
L
H
H
L
L
H
H
L
L
H
H
L
L
H
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
L = Low (Gnd)
H = High (>+3V)
* 0-7 in 3-bit mode, which is controlled by CONFIG LIMIT menu.
(See
Section 12
.) 0-65535 with 2499-DIGIO 16-bit option.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...