![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086204.webp)
2400 Series SourceMeter
®
User’s Manual
Relative and Math
8-3
Remote rel programming
Rel commands
Table 8-1
summarizes rel commands. See
Section 18
, “
CALCulate2
” subsystem, for addi-
tional information.
Rel programming example
Table 8-2
lists commands for setting up and enabling rel. These commands set up the
SourceMeter as follows:
•
Rel value: 5
•
Rel state: enabled
Table 8-1
Rel commands
Command
Description
:CALCulate2:NULL:OFFSet <n>
:CALCulate2:NULL:STATe <state>
:CALCulate2:NULL:ACQuire
Define null (rel) value (n = rel value).
Enable/disable rel (state = ON or OFF).
Automatically acquire rel value (must
have non-overflowed reading).
Table 8-2
Rel programming example
Command
Description
:CALC2:NULL:OFFS 5
:CALC2:NULL:STAT ON
Rel value =5.
Enable rel.