![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086456.webp)
2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference
18-75
Select range
RANGe <n>
:SOURce[1]:CURRent:RANGe <n>
Select range for I-Source
:SOURce[1]:VOLTage:RANGe <n>
Select range for V-Source
Parameters
2400
<n> =
-1.05 to 1.05
Specify I-Source level (amps)
-210 to 210
Specify V-Source level (volts)
DEFault
100µA range (I-Source)
20V range (V-Source)
MINimum
1µA range (I-Source)
200mV range (V-Source)
MAXimum
1A range (I-Source)
200V range (V-Source)
UP
Select next higher range
DOWN
Select next lower range
2410
<n> =
-1.05 to 1.05
Specify I-Source level (amps)
-1100 to 1100
Specify V-Source level (volts)
DEFault
100µA range (I-Source)
20V range (V-Source)
MINimum
1µA range (I-Source)
200mV range (V-Source)
MAXimum
1A range (I-Source)
1100V range (V-Source)
UP
Select next higher range
DOWN
Select next lower range
2420
<n> =
-3.15 to 3.15
Specify I-Source level (amps)
-63 to 63
Specify V-Source level (volts)
DEFault
100µA range (I-Source)
20V range (V-Source)
MINimum
10µA range (I-Source)
200mV range (V-Source)
MAXimum
3A range (I-Source)
63V range (V-Source)
UP
Select next higher range
DOWN
Select next lower range
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...