Source-Measure Concepts
6-33
Measure only (V or I)
Figure 6-17
shows the configurations for using the SourceMeter exclusively as a voltmeter
or ammeter. As shown in
Figure 6-17A
, the SourceMeter is configured to measure voltage
only by setting it to source 0A and measure voltage.
CAUTION
V-Compliance must be set to a level that is higher than the measured
voltage. Otherwise, excessive current will flow into the SourceMeter.
This current could damage the SourceMeter. Also, when connecting an
external voltage to the I-Source, set the output-off state to the high-
impedance mode. (See
Section 13
, “
Output-off states
.”)
In
Figure 6-17B
, the SourceMeter is configured to measure current-only by setting it to
source 0V and measure current. Note that in order to obtain positive (+) readings, conven-
tional current must flow from IN/OUT HI to LO.
Figure 6-17
Measure only (V or I)
V-Meter
I-Source
(0.00000
µ
A)
A. Measure Voltage Only
IN/OUT
HI
IN/OUT
LO
±
DUT (V-Source)
I-Meter
V-Source
(000.000mV)
B. Measure Current Only
IN/OUT
HI
IN/OUT
LO
DUT (I-Source)
Positive
Current
Note:
Positive current flowing out of
IN/OUT HI results in positive (+)
measurements.
Note: Use 2-wire local sensing.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...