2400 Series SourceMeter
®
User’s Manual
Source-Measure Concepts
6-43
Changing MATH function
•
If you started with only a basic measurement function selected, you can enable a
MATH function, but only the voltage, current, or resistance component of the cal-
culation will be stored in the buffer. The results of the MATH function will not be
stored.
•
If you started with a MATH function enabled, you can select a different MATH
function. The results of the new MATH function are stored in the buffer.
•
If you started with REL and/or LIMITS enabled, the buffer will stop storing read-
ings if you select a MATH function. Storage will continue if you return to the orig-
inal configuration.
See
Section 8
for more information on MATH.
Changing REL or LIMITS
•
If you started with only a basic measurement function selected, you can enable
REL and/or LIMITS, but only the voltage, current, or resistance component of the
operation will be stored in the buffer. The results of REL and/or LIMITS are not
stored.
•
If you started with a MATH function enabled, only the result of the MATH calcula-
tion will be stored in the buffer if REL and/or LIMITS is enabled.
•
If you started with REL and/or LIMITS enabled, you can change REL and or
LIMITS. The results of new REL and/or LIMITS are stored in the buffer.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...