2400 Series SourceMeter
®
User’s Manual
Ohms Measurements
4-3
Ohms measurement methods
NOTE
For the Model 2430, the following ohms measurement procedures assume that
the DC Mode of operation is selected (“Vsrc” or “Isrc” displayed in the source
field). If in the Pulse Mode (“Vpls” or “Ipls” displayed), you can select the DC
Mode by pressing CONFIG V or I, selecting the SHAPE menu item, and then
selecting DC. Pulse Mode operation is covered in
Section 5
.
There are two methods to measure ohms: auto ohms and manual ohms. When using auto
ohms, the SourceMeter operates as a conventional constant-current source ohmmeter or
DMM. To use this method, simply select an ohms measurement range (or use autorange),
and take the reading from the display. When using auto ohms, the default test current var-
ies with the ohms range, as summarized in
Table 4-1
.
NOTE
You cannot change the test current in the auto ohms mode. If you attempt to
change the source current in auto ohms, the SourceMeter will display an error
message.
With the manual ohms mode, you can select either source V or source I to make ohms
measurements, and the unit will automatically compute the resistance reading using the
V/I measurement method. After configuring the desired source and selecting a voltage or
current measuring range, select the
Ω
measurement method to display the calculated V/I
ohms reading.
NOTE
To achieve optimum accuracy, the SourceMeter measures both V and I and uses
these values in ohms calculations (with source readback enabled). The mea-
sured source value is more accurate than the programmed source value. For
remote operation, the user specifies the functions to measure. See the resistance
measurement accuracy specifications in
Appendix A
for more information.
Table 4-1
Auto ohms default test currents
Auto ohms
range
2400 and 2410
default test current
2420, 2425, 2430, and 2440
default test current
2
Ω
20
Ω
200
Ω
2k
Ω
20k
Ω
200k
Ω
2M
Ω
20M
Ω
200M
Ω
-
100mA
10mA
1mA
100
µ
A
10
µ
A
1
µ
A
1
µ
A
100nA
1A
100mA
10mA
1mA
100
µ
A
10
µ
A
10
µ
A
1
µ
A
-
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...