18-120 SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
TRIGger subsystem
The Trigger subsystem is made up of a series of commands and subsystems to configure
the Trigger Model. These commands and subsystems are summarized in
Table 18-11
.
NOTE
See
Section 11
, “
Triggering
,” for more details on triggering and the trigger
model.
Clear input triggers
CLEar
:TRIGger:CLEar
Clear pending input triggers
Description
When this action command is sent, any pending (latched) input triggers
are cleared immediately. When the SourceMeter is being triggered by
another instrument, it may inadvertently receive and latch input triggers
that do not get executed. These pending triggers could adversely affect
subsequent operation.
When using external triggering, it is recommended that TRIGger:CLEar
be sent after sending the ABORt command and at the beginning of a
program before sending an initiate command. (See “
INITiate
”
command.)
Initiate source/measure cycle
INITiate
:INITiate[:IMMediate]
Take SourceMeter out of idle state
Description
This command is used to initiate source-measure operation by taking the
SourceMeter out of idle. The :READ? and :MEASure? commands also
perform an initiation.
Note that if auto output-off is disabled (SOURce1:CLEar:AUTO OFF),
the source output must first be turned on before an initiation can be per-
formed. The :MEASure? command automatically turns the output
source on before performing the initiation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...