SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
:LATest?
:CALCulate[1]:DATA:LATest?
Read latest CALC1 result
Description
This command works exactly like CALC1:DATA? except that it returns
only the latest CALC1 result.
CALCulate2
The following commands are used to configure and control the three limit tests for DUT.
When used with a handler to provide binning operations, communication between the
SourceMeter and the handler is provided via the Digital I/O port. Many control aspects of
the digital output lines are performed from the SOURce2 Subsystem. These control
aspects include setting and clearing the digital output lines, and setting pulse width. See
“SOURce2,” page 18-100
, subsystem for details.
Select input path
FEED <name>
:CALCulate2:FEED <name>
Select input path for limit tests
Parameters
<name> =
CALCulate[1]
Use result of CALC1
VOLTage
Use measured voltage reading
CURRent
Use measured current reading
RESistance
Use measured resistance reading
Query
:FEED?
Query input path for limit tests
Description
This command is used to select the input path for the limit tests. With
CALCulate[1] selected, the specified limits will be compared to the
result of CALC1. With VOLTage selected, limits will be compared to
the voltage measurement. With CURRent or RESistance selected, limits
will be compared with the respective current or resistance measurement.
Configure and control limit tests
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...