F-18
Contact Check Function
2400 Series SourceMeter
®
User’s Manual
FAIL?
:CALCulate2:LIMit4:FAIL?
Read LIMIT 4 test result
Description
This command is used to read the results of the LIMIT 4 test:
0 = Limit 4 test passed
1 = Limit 4 test failed
Reading the results of the limit test does not clear the fail indication of
the test. A failure can be cleared by sending a :SYSTem:CLEar
command.
SOURce2 <NRf> |<NDN>
:CALCulate2:LIMit4:SOURce2 <NRf> |<NDN>
Specify pattern; LIMIT 4 failure
Parameters
<NRf> =
0 to 7 (3-bit)
Decimal value
0 to 15 (4-bit)
Decimal value
<NDN> =
0 to #b111 (3-bit)
Binary value
0 to #b1111 (4-bit)
Binary value
0 to #q7 (3-bit)
Octal value
0 to #q17 (4-bit)
Octal value
0 to #h7 (3-bit)
Hexadecimal value
0 to #hF (4-bit)
Hexadecimal value
Query
:SOURce2?
Query source value for specified limit
Description
This command is used to define the digital output fail patterns for the
Limit 4 test (0 to 7, 3-bit; 0 to 15, 4-bit).