2400 Series SourceMeter
®
User’s Manual
Limit Testing
12-25
Test parameters for this test include:
•
Source Function: current
•
Sense Function: voltage
•
Source Current: 100mA
•
Source Delay: 100ms
•
Limit 2 Upper Value: 0.85V
•
Limit 2 Lower Value: 0.75V
•
Limit 3 Upper Value: 0.82V
•
Limit 3 Lower Value: 0.78V
Figure 12-11
demonstrates graphically how parts are sorted. Diodes with a voltage
between 0.78V and 0.82V are considered good and will pass the limits test. Diodes that
test with a slightly wider voltage range are routed to QA for analysis, while those with the
largest voltage tolerance will be discarded.
Figure 12-11
Diode pass/fail limits
Table 12-3
summarizes the basic SCPI command sequence for performing a limit test for
the diode breakdown described above, and
Table 12-4
summarizes pass/fail parameters.
Note that the test connections shown in
Figure 10-8
can be used for this test.
NOTE
Additional programming steps will be necessary to test the values returned by
the :CALC2:LIM2:FAIL? and :CALC:LIM3:FAIL? queries. A returned value of
1 indicated a failure of that limit test.
Bad Diode,
Discard
Bad Diode,
to QA
Good
Diode
Bad Diode,
to QA
Bad Diode,
Discard
Low Limit 2
(0.75V)
Low Limit 3
(0.78V)
Upper Limit 3
(0.82V)
Upper Limit 2
(0.85V)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...