18-12
SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
[:SENSe[1]]
:VOLTage[:DC]
:PROTection
[:LEVel]?
Query voltage compliance limit.
✓
:TRIPped?
In voltage compliance?: 1 (yes), 0 (no).
✓
:RSYNchronize <b>
Enable or disable measure and
compliance range synchronization.
OFF
:RSYNchronize?
Query range synchronization.
:RESistance
Path to configure resistance:
✓
:MODE <name>
Select ohms mode (MANual or AUTO).
AUTO
✓
:MODE?
Query ohms mode.
:OCOMpensated <b>
Enable or disable offset-compensated
ohms.
1
OFF
✓
✓
:OCOMpensated?
Query state of offset-compensated ohms.
✓
:RANGe
Configure measurement range:
✓
[:UPPer]
<n>|UP|DOWN
Select range by specifying the expected
resistance reading.
2
2.1e5
✓
✓
[:UPPer]?
Query range.
✓
:AUTO <b>
Enable or disable auto range.
3
ON
✓
✓
:AUTO?
Query auto range.
✓
ULIMit <n>
Set auto range upper limit.
2
✓
ULIMit?
Query auto range upper limit.
✓
LLIMit <n>
Set auto range lower limit.
2
✓
LLIMit?
Query auto range lower limit.
✓
:NPLCycles <n>
Specify integration rate (in line cycles):
0.01 to 10.
4
1.0
✓
✓
:NPLCycles?
Query integration rate.
✓
:AVERage
Path to configure and control digital filter:
REPeat
✓
:TCONtrol <name>
Specify filter type (MOVing or REPeat).
5
✓
✓
:TCONtrol?
Query filter type.
✓
:COUNt <n>
Specify filter count (1 to 100).
5
10
✓
✓
:COUNt?
Query filter count.
✓
[:STATe] <b>
Enable or disable filter.
6
OFF
✓
✓
[:STATe]?
Query state of filter.
✓
1. For the Model 2430 Pulse Mode, offset-compensated ohms cannot be enabled using this command. This command is invalid (error +831).
2. The parameter values for this command are provided with the detailed command reference information. Detailed information on SCPI commands fol-
lows the last command summary table in this section.
3. For the Model 2430 Pulse Mode, auto range is invalid (error +831).
4. For the Model 2430 Pulse Mode, <n> = 0.004 to 0.100.
5. For the Model 2430 Pulse Mode, filtering is not used. This command is ignored.
6. For the Model 2430 Pulse Mode, the filter cannot be enabled. This command is invalid (error +831).
7. If auto ohms is enabled.
Table 18-6 (continued)
SENSe command summary
Command
Description
Default
parameter SCPI
Source
memory
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...