![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086362.webp)
2400 Series SourceMeter
®
User’s Manual
Status Structure
15-15
Questionable Event Register
The used bits of the Questionable Event Register (shown in
Figure 15-7
) are described as
follows:
•
Bits B0 through B7
— Not used.
•
Bit B8, Calibration Summary (Cal)
— Set bit indicates that an invalid calibration
constant was detected during the power-up sequence. This error will clear after
successful calibration of the instrument.
•
Bits B9 through B13
— Not used.
•
Bit B14, Command Warning (Warn)
— Set bit indicates that a Signal Oriented
Measurement Command parameter has been ignored.
•
Bit B15
— Not used.
Figure 15-7
Questionable event status
Questionable
Event Enable
Register
&
&
OR
To QSB of Status
Byte Register.
(See Figure 15-3.)
Warn = Command Warning
Cal = Calibration Summary
& = Logical AND
OR = Logical OR
—
(B7 - B0)
CAL
(B8)
—
(B13 - B9)
—
(B15)
Warn
(B14)
Questionable
Condition
Register
Questionable
Event Register
—
(B7 - B0)
CAL
(B8)
—
(B13 - B9)
—
(B15)
Warn
(B14)
—
(B7 - B0)
CAL
(B8)
—
(B13 - B9)
—
(B15)
Warn
(B14)
:stat:ques:cond?
:stat:ques?
:stat:ques:enab<NRf>
stat:ques:enab?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...