2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference
18-81
2410
<n> =
-1100 to 1100
Specify V-Source limit
20
Set limit to 20V
40
Set limit to 40V
100
Set limit to 100V
200
Set limit to 200V
300
Set limit to 300V
400
Set limit to 400V
500
Set limit to 500V
501 to 1100
Set limit to NONE
NONE
Set limit to 1100V
DEFault
Set limit to 1100V (NONE)
MINimum
Set limit to 20V
MAXimum
Set limit to 1100V (NONE)
2420
<n> =
-63 to 63
Specify V-Source limit
6
Set limit to 6V
12
Set limit to 12V
18
Set limit to 18V
24
Set limit to 24V
30
Set limit to 30V
36
Set limit to 36V
48
Set limit to 48V
49 to 63
Set limit to NONE
NONE
Set limit to 63V
DEFault
Set limit to 63V (NONE)
MINimum
Set limit to 6V
MAXimum
Set limit to 63V (NONE)
2425 and 2430
<n> =
-105 to 105
Specify V-Source limit
10
Set limit to 10V
20
Set limit to 20V
30
Set limit to 30V
40
Set limit to 40V
50
Set limit to 50V
60
Set limit to 60V
80
Set limit to 80V
81 to 105
Set limit to NONE
NONE
Set limit to 105V
DEFault
Set limit to 105V (NONE)
MINimum
Set limit to 10V
MAXimum
Set limit to 105V (NONE)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...