1-26
Getting Started
2400 Series SourceMeter
®
User’s Manual
Front panel control
Front panel display circuitry is controlled from the DISABLE DISPLAY configuration
menu, which is accessed by pressing CONFIG and then EDIT (or TOGGLE). To select an
option (NOW, NEVER, SWEEP, or STORE), use the EDIT cursor keys to place the cursor
on the desired option, then press ENTER.
The options for DISABLE DISPLAY are explained as follows:
NOW
— Select this option to disable the display now.
NEVER
— Select this option if you do not want the display to disable.
SWEEP
— Select this option if you want the display to disable while performing a sweep.
The display will disable as soon as sweep is started. The display will automatically re-
enable after the sweep is completed.
STORE
— Select this option if you want the display to disable when storing source-
measure readings in the buffer. The display will disable as soon as the buffer is enabled.
The display will automatically re-enable after the storage process is completed. Note that
with this option, the display will disable while performing a sweep. Sweep readings are
automatically stored in the buffer.
Remote command programming
Use the following SCPI commands to enable or disable the front panel display circuitry:
:DISPlay:ENABle OFF
Disable the display
:DISPlay:ENABle ON
Enable the display
Configuration menus
There are a number of configuration menus that can be accessed by pressing the CONFIG
key followed by the appropriate function or mode key. For example, you can configure the
voltage source by pressing CONFIG then SOURCE V. Configuration menus, which are
summarized in
Table 1-6
through
Table 1-10
, are available for the following operating
modes:
•
Measure functions (V,
Ω
, FCTN):
Table 1-6
.
•
SOURCE V, SOURCE I, and RANGE:
Table 1-7
.
•
REL, FILTER, and LIMIT:
Table 1-8
.
•
TRIG:
Table 1-9
.
•
SWEEP, DIGITS, SPEED, and STORE:
Table 1-10
.
•
ON/OFF OUTPUT and Display (EDIT or TOGGLE):
Table 1-11
.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...