2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference
18-71
Configure and control filter
NOTE
For the Model 2430 Pulse Mode, filtering is not used. Therefore, the following
filter commands are not valid for the Model 2430 Pulse Mode.
TCONtrol <name>
[:SENSe[1]]:AVERage:TCONtrol <name>
Select filter type
Parameters
<name> =
REPeat
Repeating filter
MOVing
Moving filter
Query
:TCONtrol?
Query filter type
Description
This command is used to select the type of averaging filter (REPeat or
MOVing). These filter types are explained in
Section 7
, “
Filters
.”
The number of readings that are averaged by the filter is set with the
:AVERage:COUNt command. The :AVERage:STATe command is used
to enable or disable the filter.
NOTE
For the Model 2430 Pulse Mode, this command is ignored.
COUNt <n>
[:SENSe[1]]:AVERage:COUNt <n>
Specify filter count
Parameters
<n> =
to 100
Specify filter count
DEFault
10
MINimum
1
MAXimum
100
Query
:COUNt?
Query filter count
:COUNt? DEFault
Query the *RST default filter count
:COUNt? MINimum
Query the lowest allowable filter count
:COUNt? MAXimum
Query the largest allowable filter count
Description
These commands are used to specify the filter count. In general, the fil-
ter count is the number of readings that are acquired and stored in the
filter buffer for the averaging calculation. The larger the filter count, the
more filtering that is performed.
NOTE
For the Model 2430 Pulse Mode, this command is ignored.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...