7-12
Range, Digits, Speed, and Filters
2400 Series SourceMeter
®
User’s Manual
Figure 7-3
Filter configuration menu tree
Remote filter programming
Filter commands
Table 7-4
summarizes filter commands. See
Section 18
, “
SENSe1 subsystem
,” “
Configure
and control filter
,” for more details.
Table 7-4
Filter commands
Command
Description
:SENSe:AVERage:TCONtrol <type>
:SENSe:AVERage:COUNt <n>
:SENSe:AVERage <state>
Select filter type (type = REPeat or MOVing).
Set filter count (n = count, 1 to 100).
Enable/disable filter (state = ON or OFF)*.
* For the Model 2430 Pulse Mode, the filter cannot be enabled.