![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086581.webp)
G-8
GPIB 488.1 Protocol
2400 Series SourceMeter
®
User’s Manual
Single-shot operation
Table G-5
through
Table G-7
show reading rates for measure-only, source-measure, and
source-measure-limit test single-shot operation respectively. Note that the reading rate can
be increased by >100%.
Table G-5
SCPI/488.1 reading speed comparisons
for measure-only single-shot operation (rdgs/sec)
Speed
NPLC/Trig. origin
SCPI
488.1
Improvement
Fast
0.01/internal
256.0
537.8
110.08%
Medium
0.10/internal
167.3
294.3
75.86%
Normal
1.00/internal
49.2
53.9
9.47%
Table G-6
SCPI/488.1 reading speed comparisons
for source-measure single-shot operation (rdgs/sec)
Speed
NPLC/Trig. origin
SCPI
488.1
Improvement
Fast
0.01/internal
79.5
140.0
76.15%
Medium
0.10/internal
72.9
116.9
60.27%
Normal
1.00/internal
34.9
42.3
21.20%
Note: Includes time to reprogram source to a new level before making measurement.
Table G-7
SCPI/488.1 reading speed comparisons
for source-measure-limit test single-shot operation (rdgs/sec)
Speed
NPLC/Trig. origin
SCPI
488.1
Improvement
Fast
0.01/internal
79.3
135.9
71.44%
Medium
0.10/internal
69.9
113.9
62.81%
Normal
1.00/internal
35.0
41.7
19.33%
Notes:
1. Pass/Fail test performed using one high limit and one low math limit.
2. Includes time to reprogram source to a new level before making measurement.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...