18-124 SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
TIMer <n>
:ARM[:SEQuence[1]][:LAYer[1]]:TIMer <n>
Set interval for arm layer timer
Parameters
<n> =
0.001 to 9999.999
Specify timer interval in seconds
10000.00 to 99999.99
Specify timer interval in seconds
Query
:TIMer?
Query programmed timer interval
Description
These commands are used to set the interval for the timer. Note that the
timer is in effect only if the timer is the selected control source.
DIRection <name>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:DIRection <name>
Control arm bypass
:TRIGger[:SEQuence[1]][:TCONfigure]:DIRection <name>
Control trigger bypass
Parameters
<name> =
SOURce
Enable control source bypass
ACCeptor
Disable control source bypass
Query
:DIRection?
Query state of bypass
Description
This command is used to enable (SOURce) or disable (ACCeptor) con-
trol source bypass. When enabled, operation will loop around the con-
trol source on the first pass in the layer. After that, repeat passes in the
layer are held up and will wait for the programmed control source event.
INPut <event list>
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:INPut <event list> Enable event detectors
Parameters
<event list> = SOURce
Enable Source Event Detector
DELay
Enable Delay Event Detector
SENSe
Enable Measure Event Detector
NONE
Disable all event detectors in Trigger
Layer
NOTE
Each event in the list must be separated by a comma (i.e. trigger:input source,
delay, sense).
Query
:INPut?
Query enabled event detectors in Trigger
Layer
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...