2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference
18-5
:CALCulate2
:LIMit3
Path to control LIMIT 3 test:
✓
:UPPer
Configure upper limit:
✓
[:DATA] <n>
Specify upper limit (-9.999999e20 to
9.999999e20).
1
✓
✓
[:DATA]?
Query upper limit.
✓
:SOURce2 <NRf>
| <NDN>
Specify output fail pattern (0 to 7,
3-bit; 0 to 15, 4-bit).
15
✓
:SOURce2?
Query fail bit pattern.*
:LOWer
Configure lower limit:
✓
[:DATA] <n>
Specify lower limit (-9.999999e20 to
9.999999e20).
-1
✓
✓
[:DATA]?
Query lower limit.
✓
:SOURce2 <NRf>
| <NDN>
Specify output fail pattern for grading
mode. (0 to 7, 3-bit; 0 to 15, 4-bit).
15
✓
:SOURce2?
Query fail bit pattern.*
:PASS
Path to specify pass pattern for sorting
mode:
:SOURce2 <NRf>
| <NDN>
Specify output pass pattern (0 to 7,
3-bit; 0 to 15, 4-bit).
15
✓
✓
:SOURce2?
Query pass bit pattern.*
:STATe <b>
Enable or disable Limit 3 test.
OFF
✓
✓
:STATe?
Query state of Limit 3 test.
✓
:FAIL?
Return result of Limit 3 test: 0 (pass) or
1 (fail).
✓
:LIMit4
Path for LIMIT 4 test (contact check option
only).
:SOURce2 <NRf>
| <NDN>
Specify output fail pattern (0 to 7, 3-bit;
0 to 15, 4-bit).
15
:SOURce2?
Query fail bit pattern.*
:STATe <b>
Enable or disable Limit 4 test.
OFF
✓
✓
:STATe?
Query state of Limit 4 test.
✓
:FAIL?
Return result of Limit 4 test: 0 (pass) or
1 (fail).
✓
Table 18-1 (continued)
CALCulate command summary
Command
Description
Default
parameter SCPI
Source
memory
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...