![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086373.webp)
16-6
Common Commands
2400 Series SourceMeter
®
User’s Manual
*TRG programming example
The command sequence in
Table 16-4
configures the SourceMeter to be controlled by bus
triggers. The last command, which sends a bus trigger, triggers one measurement. Each
subsequent bus trigger will also trigger a single measurement.
NOTE
With :ARM:SOURce BUS selected, do not send any commands (except *TRG,
GET, DCL, SDC, IFC, and ABORt) while performing source-measure opera-
tions. Otherwise, erratic operation will occur.
*TST? — self-test query
Use this query command to perform a checksum test on ROM. The command places the
coded result (0 or 1) in the Output Queue. When the SourceMeter is addressed to talk, the
coded result is sent from the Output Queue to the computer.
A returned value of zero (0) indicates that the test passed, and a value of one (1) indicates
that the test failed.
Table 16-4
*TRG programming example
Command
Description
*RST
:ARM:SOUR BUS
:ARM:COUN INF
:OUTP ON
:INIT
*TRG
Restore GPIB defaults.
Select BUS control source.
Set arm layer count to infinite.
Turn on output.
Take SourceMeter out of idle.
Trigger one measurement.
Run self test and read result
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...