![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086517.webp)
A-10
Specifications
2400 Series SourceMeter
®
User’s Manual
Case IV: Auto-Zero disabled and all measurements disabled
Figure
A
-4
Case IV timing diagram
Trigger Latency:
225µsec max
Source Configuration: 50µsec
max
Firmware Overhead:
310µsec for Source V
590µsec for Source I
Source On Time
≅
Source Configu Source Delay + Firmware Overhead
Example:
Source Delay = 0
Source On Time
≅
50µsec + 0 + 125µsec
≅
360µsec for Source V
≅
640µsec for Source I
Trigger
Event
Source
On
Source On Time
Source
Off
Trigger
Delay
Trigger
Latency
Source
Configuration
Source
Delay